  Salt spray chamber price of the four failures of analysis, salt spray chamber as one of "three" test equipment, product research and development in many electronic test equipment fully play a critical effect. Summary of the difficult problems to deal with at the price of salt spray chamber to do a detailed explanation, pictures with text, there is a certain expectation of customer assistance.

  First, the actual salt spray chamber did not operate the dashboard lights lit Flag:

  Regional response to this problem, one should first check the concept of environmental chambers 2 plus pure water, a saline plus test area, it is not added enough water, because the water level is too low these two, salt spray chamber He is unable to work.

  That is followed by the actual operation of the instrument panel will not cause abnormal work (this problem usually occurs very little), the solution that is removable or troubleshooting dashboard.

  Second, the aqueous salt spray chamber show less long bright lights:

  This problem has many customers are not very grasp, careful interpretation of what people give everybody: the salt spray chamber with three operating panel indicator lights less water, less water two indicator lights, a low salt indicator lights, if these three a display lights were on, laboratory tests were meant less water, less saturated buckets of water, saline test is not enough. This time you are full of people only have to water in the aqueous phase matching at least three (added to the three display lights blew that is already filled with oil) salt spray chamber can all normal work. Must be noted that, turn on the water at saturation barrels a gate valve manipulation, the need to turn on the water when the valve is turned on, be sure to turn off the valve after the water is full of oil! Barrels of saturation inside the water fountain plaza will be like or do not turn off the valve when do sprayer test Like splash out.

  Customers have been added brine then added under low salt indicator lamp has not blown out, that is curing this problem bottom end of the reservoir housing salt brine stored inside a rubber stopper is not blocked, causing all the brine is injected into the lab detection inside. Most of the problem leading to the client that is the reason that done washing machine salt spray test, the vulcanized rubber forget to pull out the stopper will be back in the plug row brine. So customers clean up after the salt spray test chamber must remember the simplicity of homing traffic fortress!

  Third, do not salt spray chamber sprayer:

  When you do the test, press and hold the power switch sprayers, salt spray chamber does not spray, some customers have encountered. That is the reason there are problems arise salt spray chamber and nozzle clogging. Since the nozzle is a high-precision equipment, the only straight through pores mm. If people test tap water inside a residue, very easy to block the nozzle, so people generally advocate customers if there is a standard, then have to apply a test of pure water.

  That is another reason inadequate salt spray chamber, nebulizer operating pressure, this time people only have to pull out the pressure relief valve, adjusted to the required working pressure can be.

  That is also the reason will fall into the intake manifold, into the housing without air, that is to find solutions to the respiratory tract fall, connected, fastened with a strap.

  Fourth, the salt spray chamber "steam leakage"?

  Before people received a customer reflect say the bottom of a small amount of salt spray chamber salt spray vapor went out, people send professional and technical personnel in the past queries, customers get that done originally test of time cleaning machine, accidentally shell the coupling member inside the intake manifold vulcanized rubber plug to get off, and there is overflow salt spray vapor. After washing machine so people must ensure that each race vulcanized rubber plugs are tight. Since this is all some small problems, pay attention to the point when the customer can prevent the actual operation.